C-Link Protocol Commands
Measurements
B-8
Model 48
i
Instruction Manual
Thermo Electron Corporation
Measurements
avg time
high avg time
low avg time
These commands report the averaging time in seconds when operating in
single range, or averaging time used with the high and low ranges when
operating in dual or auto range mode. The example below shows that the
averaging time is 300 seconds, according to Table B-2.
Send:
avg time
Receive:
avg time 11:300 sec
set avg time
selection
set high avg time
selection
set low avg time
selection
These commands set the averaging time, high and low averaging times,
according to Table B-2. The example below sets the low range averaging
time to 120 seconds.
Send:
set low avg time 8
Receive:
set low avg time 8 ok
stream time
Reports/sets a time stamp to streaming data or not
B-22
temp comp
Reports/sets temperature compensation on or off
B-29
time
Reports/sets current time (24-hour time)
B-31
up
Simulates pressing up pushbutton
B-25
zero
Sets zero/span valves to zero mode
B-29
Table B-1.
C-Link Protocol Commands
Command
Description
Page
Table B-2.
Averaging Times
Selection
Averaging Time (seconds)
0
1 seconds
1
2
2
5
3
10
4
20
5
30
6
60
7
90
8
120
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part number 101891 00 8Apr2005 ...
Страница 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Страница 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 236: ...System Description I O Components 8 14 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Страница 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...
Страница 243: ...Thermo Electron Corporation Model 48i Instruction Manual 9 7 Figure 9 5 Retrofit Rack Mounting ...
Страница 244: ...Optional Equipment Mounting Options 9 8 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 292: ...C Link Protocol Commands Record Layout Definition B 46 Model 48i Instruction Manual Thermo Electron Corporation ...