Operation
Alarms Menu
Thermo Electron Corporation
Model 48
i
Instruction Manual
3-93
Sample Flow
The Sample Flow screen displays the current sample flow reading and set
the minimum and maximum alarm limits. Acceptable alarm limits range
from .350 to 1.500 LPM. If the sample flow reading goes beyond either the
minimum or maximum limit, an alarm is activated. The word “ALARM”
appears in the Run screen and in the Main Menu.
•
In the Main Menu, choose Alarms >
Sample Flow
.
•
Use
and
to move the cursor up and down.
•
Press
to select a choice.
•
Press
to return to the Alarms menu or
to return to the
Run screen.
Min and Max Sample Flow Limits
The Minimum Sample Flow alarm limit screen is used to change the
minimum sample flow alarm limit. The minimum and maximum sample
flow screens function the same way.
•
In the Main Menu, choose Alarms > Sample Flow >
Min
or
Max
.
•
Use
and
to increment or decrement the numeric value.
•
Press
to save set value as actual vlaue.
•
Press
to return to the Sample Flow menu or
to return to
the Run screen.
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
SAMPLE FLOW:
ACTUAL 0.750 LPM
>MIN 0.350 LPM
MAX 1.500 LPM
RANGE AVG DIAGS ALARM
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part number 101891 00 8Apr2005 ...
Страница 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Страница 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 236: ...System Description I O Components 8 14 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Страница 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...
Страница 243: ...Thermo Electron Corporation Model 48i Instruction Manual 9 7 Figure 9 5 Retrofit Rack Mounting ...
Страница 244: ...Optional Equipment Mounting Options 9 8 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 292: ...C Link Protocol Commands Record Layout Definition B 46 Model 48i Instruction Manual Thermo Electron Corporation ...