4-1
Chapter 4
Calibration
This chapter describes the procedures for performing a standard zero/span
calibration and Multi-Point calibrations of the Model 48
i
. The information
described here is more than adequate to perform the calibration. However, if
greater detail is needed, please refer to the
Quality Assurance Handbook for
Air Pollution Measurement Systems
1
.
The following sections discuss the required apparatus and procedure for
calibrating the instrument.
Equipment Required
The following equipment is required to calibrate the instrument:
CO Concentration
Standard
A cylinder of CO in air containing an appropriate concentration of CO
suitable for the selected operating range of the analyzer under calibration is
necessary. The assay of the cylinder must be traceable either to a National
Institute of Standards and Technology (NIST) CO in Air Standard
Reference Material (SRM) or an NIST/EPA approved gas manufacturer's
Certified Reference Material (CRM).
A recommended protocol for certifying CO gas cylinders against a CO,
SRM or CRM is given in the
Quality Assurance Handbook
1
. The CO gas
cylinder should be recertified on a regular basis determined by the local
quality control program.
Zero Air Generator
Calibration requires zero air that is free of contaminants which will cause a
detectable response on the CO analyzer. The zero air should contain <0.01
ppm CO. Since the Model 48
i
is virtually interference free, it is only
necessary to ensure that CO has been removed.
It should be noted that zero air as supplied in cylinders from commercial
suppliers typically contains CO concentrations in the 0.1 - 0.3 ppm range.
So cylinder zero air should be scrubbed of the residual CO prior to its use in
the Model 48
i
as a dilution gas or a zero standard. Room air which has been
scrubbed of CO can also be used as the zero air source.
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part number 101891 00 8Apr2005 ...
Страница 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Страница 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 236: ...System Description I O Components 8 14 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Страница 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...
Страница 243: ...Thermo Electron Corporation Model 48i Instruction Manual 9 7 Figure 9 5 Retrofit Rack Mounting ...
Страница 244: ...Optional Equipment Mounting Options 9 8 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 292: ...C Link Protocol Commands Record Layout Definition B 46 Model 48i Instruction Manual Thermo Electron Corporation ...