Operation
Alarms Menu
Thermo Electron Corporation
Model 48
i
Instruction Manual
3-91
Min and Max Bench Temperature
Limits
The Minimum Bench Temperature alarm limit screen is used to change the
minimum bench temperature alarm limit. The minimum and maximum
Bench Temperature screens function the same way.
•
In the Main Menu, choose Alarms > Bench Temp >
Min
or
Max
.
•
Use
and
to increment or decrement the numeric value.
•
Press
to save set value as actual vlaue.
•
Press
to return to the Bench Temperatures menu or
to
return to the Run screen.
Pressure
The Pressure screen displays the current reaction chamber pressure reading
and set the minimum and maximum alarm limits. Acceptable alarm limits
range from 250 to 1000 mmHg. If the pressure reading goes beyond either
the minimum or maximum limit, an alarm is activated. The word
“ALARM” appears in the Run screen and in the Main Menu.
•
In the Main Menu, choose Alarms >
Pressure
.
•
Use the
and
to move the cursor up and down.
•
Press
to select a choice.
•
Press
to return to the Alarms menu or
to return to the
Run screen.
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
x
BENCH TEMPERATURE:
ACTUAL MIN: 40.0
o
C
SET MIN TO: 44.0
o
C ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part number 101891 00 8Apr2005 ...
Страница 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Страница 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 236: ...System Description I O Components 8 14 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Страница 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...
Страница 243: ...Thermo Electron Corporation Model 48i Instruction Manual 9 7 Figure 9 5 Retrofit Rack Mounting ...
Страница 244: ...Optional Equipment Mounting Options 9 8 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 292: ...C Link Protocol Commands Record Layout Definition B 46 Model 48i Instruction Manual Thermo Electron Corporation ...