Model 3010 Incubator _______________________________________________________________________Remote Alarm
5 - 1
Section 5 - Remote Alarm
5.1 Connecting the Remote Alarm Contacts
A set of relay contacts are provided to monitor alarms
through an RJ-11 telephone style connector on the back of the
cabinet. Refer to Figure 1-13 for the location of the alarm con-
nector.
The remote alarm provides a NO (normally open) output, a
NC (normally closed) output and COM (common). Refer to
Figure 5-1.
The contacts will trip on a power outage, temperature
alarm or CO
2
alarm. See Table 3-1.
Figure 5-1