1
Site Preparation
Power Requirements
Thermo Scientific
TSQ 8000 Evo Preinstallation Requirements Guide
9
• Power must be single-phase.
• Wall outlets must have earth-ground hard-wired to the main panel.
• Included power cords are 2 m (6 ft) long .
Contact your local Customer Service office to discuss power cordset concerns.
3. The TSQ 8000 Evo MS comes with the required number and type of power cords for
your region.
will help you identify the correct power cord for your region and
instrument. If you need to replace your power cord or buy an additional cord, you may
use any brand of power cord, as long as it is appropriate for your region.
Table 4.
Power cord identification information by region. (Sheet 1 of 2)
Region
C13 Power Cords (For PC, Monitors,
and Autosamplers)
C19 Power Cords (For GC
and MS)
North American 120V
North American 250V
Japanese 125V
Switzerland 250V
Australian 250V
China 250V