IMPORTANT PREINSTALLATION INFORMATION... PLEASE READ
TSQ 8000 Evo
Installation Request Form
Dear Customer:
Read the
TSQ 8000 Evo Preinstallation Requirements Guide
, and then print and complete the following installation request form.
After all items on the form are fulfilled, sign and date the form. Then, scan and email, mail, or fax this form to your local
Thermo Fisher Scientific sales/service office. The address and fax number for your local office are located on the following pages.
Print your name, company name, and company address clearly below:
Name ________________________________________________________
Company _____________________________________________________ Telephone____________________
Address ___________________________________________________________________________________
City ___________________________________ State ____________ Country ___________________
Signature ________________________________________________ Date _____________________
Principal Operator Level of Experience
GC, Injector and Column Knowledge:
Experienced
Moderate
Limited
Mass Spectrometer Theory Knowledge:
Experienced
Moderate
Limited
TraceFinder
Data System Knowledge:
Experienced
Moderate
Limited
Additional Information
Have any special acceptance specifications been agreed to in the contract?
Yes
No
If
YES
, attach full details of specifications.
Is there any additional equipment that needs to be interfaced to the system?
Yes
No
If
YES
, attach full details of additional equipment.
Note
We reserve the right to invoice you for the Field Service Engineer’s time if the installation requirements are
not met on the date of the installation. To avoid any additional cost, please ensure your site is properly prepared.
1. All laboratory remodeling has been completed.
2. Your
TSQ 8000 Evo
system is on site.
3. The principal operator is scheduled to be available
during the installation / certification period.
4. Doorways, hallways, and so on are a minimum width of
81 cm (32 in.).
5. Available floor area is sufficient and flooring will
support the load.
6. Sufficient bench space is available for all of the
equipment. List the following:
Width: ________________________________
Depth: ________________________________
Height: ________________________________
7. Workbench can support the load of the system and is
free from vibration.
8. Lighting is adequate.
9. Work area is free of magnetic disruption and
electrostatic discharge.
10. Main power is installed and is in compliance with local
electrical codes.
11. Power for test and cleaning equipment is installed.
12. Match power cables specified on the sales order. Refer
to the power requirements for more information.
13. Voltage of power outlet has been measured. Note
measured
voltage: ________________________
14. Air conditioning is adequate for temperature, humidity,
and particulate matter control. The laboratory can be
maintained at a constant temperature, between 15 and
35 °C (59 and 95 °F).
15. Relative humidity is between 5% and 95% with no
condensation.
16. All gases required are on site, gas lines are installed, and
gas regulators that terminate in 1/8” fittings are
available.
List gases and purity: ______________________
17. All relevant safety regulations have been followed.
18. I have read and agree to the End User License
Agreements for any software purchased with my system:
Signature: