44
Warning
Do not attempt to open sterilizer
door until the display reads "Cycle
Complete". Pressure within the
chamber could cause the door to
open with extreme force, possibly
causing personal injury.
Reprocess your load in the event
the sterilizing cycle has been ter-
minated prematurely. The load
may not be sterile when the steril-
izing cycle has been terminated
prematurely.
Failure to drain excess water from
the sterilizer chamber after a cycle
abort may result in subsequent
unsterile loads as a result of con-
tact with water in the chamber.
Always drain excess water from
the chamber after aborting a cycle.
Allow chamber to cool before
draining excess water.
Note
If display shows: "The Mechanical
OTP has Tripped. Shutoff Unit
Power and Reset - Caution - Unit
Hot!!" Reset as described under
"Resetting the Mechanical OTP
Sensor" in Service Manual.
Emergency Off (Cycle Abort)
Pressing the "Stop" button will immediately terminate steriliza-
tion. Allow the sterilizer to cool and press the "Stop" button
again to reset the sterilizer. Drain excess water from the ster-
ilizer chamber. Proceed with a new sterilization cycle normal-
ly.
When terminating a load, pressing the "Stop" button will
vent steam from the chamber back into the reservoir.
Resetting the Automatic OTP
Display shows: "Cycle Aborted Due to AUTO OTP Device.
Press Stop Shutoff Unit Power to Reset."
1.
Press the "Stop" button.
2.
Wait for the display to read "Select Cycle".
3.
Turn Power Switch off and then on to reset the
Low Water Cutoff.
4.
Open the chamber door and allow the chamber to
cool.
5.
Select your desired cycle and run the cycle with a
loaded chamber to verify operation.
6.
In the event of a repetition of fault call Customer
Service.
Special Operating Conditions