![Thermo Scientific ST75920-33 Скачать руководство пользователя страница 41](http://html.mh-extra.com/html/thermo-scientific/st75920-33/st75920-33_operation-manual_1104821041.webp)
41
Printer Tape Description
The printer tape is designed to be easily read as a cycle is
running, allowing you to exam at a glance the progress of the
current cycle. As a result, a complete tape must be read from
the bottom up. (Sample on page 41)
When you turn on the sterilizer, the printer will print
"Barnstead|Thermolyne Sterilemax" and the software version
number. This will print only immediately after the sterilizer has
been turned on.
When you press one of the preset cycle buttons, the print-
er will print the name of the cycle selected, the parameters
set for that cycle and the selected pressure units. If you
choose a default optional cycle, the printer will print the cur-
rent default parameters. If you bypass the default optional
cycle for a standard, the printer will print only the name of the
cycle and the selected pressure units, since the parameters
have not been set.
When you press "Start Cycle", the printer will print the
number of cycles that have been run since the unit was
turned on, the current date and time, the notation "Cycle
Started" and column headers.
During the cycle, the printer will print the elapsed time, the
current chamber temperature and the current chamber pres-
sure at one minute intervals, along with a status notation.
When you press "Start Cycle", the printer will print the num-
ber of cycles that have been run since the unit was turned on,
the current date and time, the notation "Cycle Started" and
column headers.
During the cycle, the printer will print the elapsed time, the
current chamber temperature and the current chamber pres-
sure at one minute intervals, along with a status notation.
When the sterilizer chamber achieves the temperature set
by the cycle parameters, the printer will immediately print a
record line, regardless of the time since the last record was
printed. The status notation will change from "Heatup" to
"Exp", indicating that the load is being exposed to the temper-
ature set forth in the cycle parameters. The printer will contin-
ue to print record lines at one minute increments, starting one
minute after the record line printed at the start of the expo-
sure phase but still showing the total elapsed time for the
cycle.
Default Optional Cycle Printout
P
RINTER