![Thermo Scientific ST75920-33 Скачать руководство пользователя страница 35](http://html.mh-extra.com/html/thermo-scientific/st75920-33/st75920-33_operation-manual_1104821035.webp)
35
R
ECOMMENDED
M
ONITORING
Packs
Textiles
Prepare a pack of 14 towels (per AAMI ST-8). Place three
biological indicators in the center of the pack, between the
sixth and seventh towels. Wrap the pack with two strips of
autoclave tape, placing the tape around the width of the
pack so that the two strips divide the length of the pack
into thirds.
Pouches
Place a biological indicator together with a surgical instru-
ment into a self-sealing pouch. Place the pouch with the
indicator in the middle of the tray. Place four other surgical
instruments into pouches and place them on the tray.
Place the tray in the center rack of the sterilizer chamber.
Optional Cycle: For Dental Handpieces Only
Place a biological indicator together with one dental hand-
piece in a self-sealing pouch. Place the closed pouch in
the front of the sterilizer tray. Place the tray in the lowest
level of the chamber rack.