Sterilizer Features
Optional Printer
The optional printer available is an impact printer that records
each cycle's performance. See page 39 for more information.
Pressure Monitor
The pressure monitor indicates the pressure inside the chamber.
This monitor's readings can be found on the display as each
cycle is running.
Door Security
While the chamber is under pressure, the door is held securely
closed with three separate locking systems. First, there is a lock-
ing handle that you must tighten to hold the door fully closed. The
locking handle turns clockwise to close and counterclockwise to
open. Second, there is a pressure actuated locking pin which
automatically secures the door, preventing it from being inadver-
tently opened while there is pressure in the chamber. Finally,
there is a backup catch which will prevent the door from opening
fully even if there is a small amount of pressure in the chamber.
You release this catch with a lever which is located in the lower
left side of the door. To fully open the door, pull the door release
lever toward you.
Optional Non-Recirculating Tank
The optional non-recirculating tank is used to collect the remain-
ing cycle water after a cycle has completed. This prevents any
debris from the cycle to go back into the fresh water reservoir.
Door Switch
Actuated by the door, the door switch prevents operation of the
sterilizer if the door is not fully closed.
Pressure Relief Valve
This valve protects the coded vessel from pressures exceeding
the pressure rating of the vessel. The valve opens and releases
steam in the event that pressure in the vessel exceeds 45 PSIG
(310 kPa). The pressure line is located in the back of the unit. An