The optional printer provides a paper tape record for future
reference and verifies sterilizer performance, load control
identification and quality assurance checks for each cycle.
During a sterilization cycle, the printer prints a record of the
chamber temperature, pressure and status of the cycle at 1-
minute intervals.
Connecting and Starting the
Printer
1.
Make sure the printer and sterilizer are
turned OFF.
2.
Connect the enclosed data cable to the
RS232 connector on the back of the steriliz-
er and the power cord to the power supply.
3.
Turn the printer ON.
4.
Turn the sterilizer ON. The printer should
print "Barnstead/Thermolyne Sterilemax,"
"Software Version X.XX.XX"
5.
If the printer does not print heading, verify
the "Printer Mode" is set to used. See
"Changing Printer Mode."
6.
Turn the sterilizer off and then back on to
verify the printer prints out
"Barnstead/Thermolyne Sterilemax,"
"Software Version X.XX.XX"
7.
The printer is now ready to print.
Control Panel (See Figure B)
Power Switch
Located on the front of the printer, this switch turns power to
the printer ON and OFF. (See Figure A.) The green POWER
light will illuminate when the printer is ON.
Optional Printer
Figure A: Printer
ON/OFF
Button
Figure B: Control Panel
Paper Feed Button
Paper Out Light
Error Light
Power Light