Warning
Use caution when opening the
sterilizer door. Stand to the side
behind the door and open door
slowly and partially to protect
yourself from escaping steam or
boiling liquids and exploding
flasks.
Do not attempt to open sterilizer
door until the display reads "Cycle
Complete." Pressure within the
chamber could cause the door to
open with extreme force, possibly
causing personal injury.
O
PERATION
Warning
Processing goods at less than rec-
ommended time/temperature
could result in unsterile goods.
Setting Drying Time
(For Pack, Wrapped and Unwrapped cycles only.)
If drying is desired, set the drying time when you select
the sterilization cycle. After you select a Pack, Wrapped or
Unwrapped cycle, the cycle parameters will be displayed.
After the parameters have been displayed, you will be
prompted to enter a figure for drying time. You will find that
the default time — displayed at the prompt — is the opti-
mal setting for most loads.
If your application or load composition requires a differ-
ent drying time to achieve appropriate drying, you can
change the drying time setting when the display prompts
you. Use the up or down arrow to adjust the displayed dry-
ing time until your desired drying time is displayed.
Press "Start Cycle"
Processing will run automatically after you press "Start
Cycle."
Drying Sterilized Loads
At the completion of the sterilization cycle, turn the lock-
ing handle counter clockwise until the door is held in place
by the backup catch. The door must be opened approxi-
mately 1 inch to operate the drying cycle.
To prevent con-
tamination of wet packs and wrapped items, do not
open the door more than 1 inch during the drying
cycle.
Press "Start Cycle" to start the drying cycle.
When you press "Start Cycle," the Drying Indicator will
be displayed, the chamber will heat to a low temperature
and the selected drying time will begin to count down. At
the end of the drying cycle, a five second alarm will sound.
Optional Drying Cycles
If you desire to run a drying cycle without running a
sterilizing cycle with it, you may. Press the "Optional
Cycle" button and scroll the setpoint temperature below
100°C. You will enter a menu for an optional drying cycle.
Set the drying time as instructed above, then press "Start