Altitude Adjustment
For proper operation, the sterilizer program accounts for the
altitude in controlling certain functions. The sterilizer is
shipped with a factory-set altitude value suitable for all alti-
tudes up to an altitude of 1000 feet above sea level. If your
altitude is above 1000 feet above sea level, you will need to
enter a new value for the altitude adjustment. To change the
altitude adjustment value:
1.
At the "Select Cycle" screen, press the "Select"
key and then the up arrow key. The sterilizer will
display a "Configuration Info" screen, then it will
automatically change to displaying the "Change
Time" screen.
2.
At the "Change Time" screen, press the "Select"
key and then the "Liquids" key.
3.
At the "Altitude Adjustment" screen, press the up
arrow to select "Yes".
4.
At the "Current Altitude" screen, press the up and
down arrows to set the value for your location's
altitude to the nearest one hundred feet.
5.
After you have set your altitude value, press the
"Start" key to return to the "Select Cycle" screen.