Step
Touchscreen operation
Screen displays and notes
2
Enter the run parameters for
step 1 as described in "(1)
Registration Procedures for
Step-Mode Runs."
Registering a rotor will make it possible to compute and
display RCFmax and RCFavg values from the set SPEED
value.
3
Press the
[Next Step]
button
and enter the run
parameters for step 2 from
the run parameter
entry/display screen.
When a SPEED of 0 rpm is set, TIME is automatically set to
"HOLD."
An arbitrary TIME setting cannot be made.
4
Press the
[Next Step]
button
and enter the run
parameters for step 5 from
step 3.
5
Finally, press the
[Register]
button.
The entered values are displayed in the STEP 1 to 5 fields.
To correct entry errors, press the STEP field you want to
correct, and press the
[Edit]
button to make corrections
when the field turns blue.
3-3-1-3 Step-Mode Operation Procedures
STEP 2
SPEED: 0 rpm
TEMP
:
15.0
°
C
[Next Step]
button
●
The run parameter
entry/display screen
appears.
[Next Step]
button
STEP 1
SPEED: 50,000 rpm
TIME
:
1 hour
TEMP
:
15.0
°
C
ACCEL: 9
DECEL: 7
STEP No.
[Register]
b
●
The run parameter
entry/display screen
in step 2 appears.
●
The run parameter
setting screen
reappears.
[X]
button
3-46