(3) Procedures for Registering Step-Mode Operation that Includes Stops
Follow the instructions in "(1) Registration Procedures for Step-Mode Operation." Perform the
registration procedures described below.
[Setting example]
The following describes the setup procedures for continuous operation under the five run
parameters including a 0 rpm settings shown in the figure below.
The operations described in Fig. 3-3-1 (2), if performed according to this example, will be
performed automatically.
Step 1
Step 2
Step 3
Step 4
Step 5
Speed
50,000 rpm
0 rpm
50,000 rpm
0 rpm
50,000 rpm
Run time
1 hour
HOLD
1 hour
HOLD
1 hour
Temperature
15.0
°
C
15.0
°
C
15.0
°
C
15.0
°
C
15.0
°
C
Acceleration
mode
9
-
9
-
9
Deceleration
mode
7
-
7
-
7
Step
Touchscreen operation
Screen displays and notes
1
As described in steps 1 to 8
in "(1) Procedures for
registration of run
parameters" in Section
3-3-1-1 "Program
registration procedures,"
enter the Program No.,
registered rotor, user and
remarks, select a folder color
and press the
[Next]
button.
In this example, the steps
are stored in program No.
11, blue folder.
Fig. 3-3-1 (2) Example of Step-Mode Operations that include stops
3-3-1-3 Step-Mode Operation Procedures
Program registration
screen
[Edit]
button
Run parameter setting/display
screen
Step 1
1 hour
50,000
Speed
(rpm)
Step 2
The door
can be
opened
0
1 hour
1 hour
Step 3
Step 5
Step 4
The door
can be
opened
End of run
[Next]
button
3-45