Methods for setting run parameters are described on the next page.
Note 1) If you enter the wrong value, press the
[CE]
button and enter the correct value.
If you have already pressed the
[Enter]
button, repeat the process and enter the
correct value.
2) When more than one SPEED, TIME, or TEMP run parameter is selected, you do not
need to press the
[Enter]
button. When you touch the field of another run parameter,
the currently entered item is confirmed.
3) To set continuous run (HOLD) and specify the time during operation, enter the sum of
time elapsed so far plus the desired remaining run time.
For example, if the centrifuge has been running for 5 hours and you want operation to
stop 1 hour and 30 minutes later, touch the TIME field and enter
[6][ : / . ] [3] [0]
.
3-2-1 Setting Run Parameters