3-3-4 RTC (Real Time Control) Operation
This centrifuge contains an internal clock allowing you to operate the centrifuge between a set start
time and end time. This function for operating the centrifuge between set time intervals is called Real
Time Control (RTC).
The RTC operation procedures are described using examples below.
Example: To operate the centrifuge under the following run parameters starting from rotor installation
on October 24 and removing the samples around 9:30 the next morning.
(1) Rotor: T-8100
(2) Speed: 100,000 rpm
(3) Separation time: 3 hours
(4) Control temperature: 4
°
C
(5) Acceleration mode: 9
(6) Deceleration mode: 9
In this example, set above run parameters (1) to (6), then set the RTC operation start time to 6:30
October 25 and start the centrifuge.
(Note that setting an end time of 9:30 instead of the start time (6:30) will perform the same
operation.)
START button
ON
3-3-4 RTC (Real Time Control) Operation
April 1 PM 4:30
Around 9:34
Spe
ed
(r
pm)
Auto start
Fig. 3-3-4 RTC Operation Example
End time
Stop
AM 6:30
[START]
button
October 24
October 25
3 h
3-54