Introduction and Description
Technical Data
Thermo Fisher Scientific
Sorvall
ST 40
1-3
Technical Data
The technical data of the
Sorvall
ST 40 is listed in the following table.
Table 1-1.
Technical Data
Sorvall
ST 40
Feature
Value
Environmental conditions
- use in interior spaces
- altitudes of up to 2,000 m above sea level
- max. relative humidity 80% up to 31 °C; decreasing linearly up to 50%
relative humidity at 40 °C.
permissible ambient temperature
+2° C to +35° C
Overvoltage category
II
Pollution degree
2
Heat dissipation
ventilated
refrigerated
230V
120V
230V
120V
4600 BTH/h
4700 BTH/h
5100 BTH/h
6300 BTH/h
IP
20
running time
unlimited
maximum speed n
max
15200 rpm (depending on the rotor)
minimum speed n
max
300 rpm
maximum RCF value at n
max
25311 xg 25314
maximum kinetic energy
< 62,5 kJ
noise level at maximum speed
< 66 dB (A)
temperature setting range
-10 °C to +40 °C
Dimensions
ventilated
refrigerated
Height
Width
Depth
Table top height
360 mm
550 mm
670 mm
310 mm
360 mm
745 mm
670 mm
310 mm
weight without rotor
86 kg
116 kg
testing standards
- all devices manufactured and tested in compliance with:
- IEC 61010-1 2
nd
Edition
- IEC 61010-2-020 2
nd
Edition
- Pollution degree 2,
- Overvoltage category II
Содержание Sorvall ST40
Страница 1: ...50119928 1 Thermo Fisher Scientific Sorvall ST 40 Sorvall ST 40R Instruction manual ...
Страница 2: ......
Страница 3: ...PN 50119928 1 Print 03 09 Instruction manual Sorvall ST 40 Sorvall ST 40R ...
Страница 6: ......
Страница 10: ...Tables iv Sorvall ST 40 Thermo Fisher Scientific ...
Страница 12: ...Figures vi Sorvall ST 40 Thermo Fisher Scientific ...
Страница 20: ...xiv Sorvall ST 40 Thermo Fisher Scientific ...
Страница 28: ...Introduction and Description Rotor selection 1 8 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 36: ...Before use Storage 2 8 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 37: ...Thermo Fisher Scientific Sorvall ST 40 3 1 Chapter 3 Control panel Control panel on page 3 2 Keys on page 3 4 ...
Страница 58: ...Operation Switching the centrifuge off 4 18 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 72: ...Maintenance and care Service of Thermo Fisher Scientific 6 12 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 80: ...Troubleshooting When to contact customer service 7 8 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 88: ...Warranty B 2 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 90: ...Index 2 Sorvall ST 40 Thermo Fisher Scientific Transporting the centrifuge 2 3 W WEEE Conformity 1 1 ...
Страница 93: ......