Troubleshooting
Troubleshooting by user
7-6
Sorvall
ST 40
Thermo Fisher Scientific
E-30
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-33
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Overpressure in the refrigeration unit.
Clean the air inlet for the condenser.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-34-E-36
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-40
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
The centrifuge accelerates too slowly.
Is the rotor properly installed?
Check wether you selected the right bucket.
Is it easy to turn the rotor when the lid is open?
Does the rotor rub against the device?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-41-E-74
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Failure message
Problem with centrifuge
Possible causes and cures
Содержание Sorvall ST40
Страница 1: ...50119928 1 Thermo Fisher Scientific Sorvall ST 40 Sorvall ST 40R Instruction manual ...
Страница 2: ......
Страница 3: ...PN 50119928 1 Print 03 09 Instruction manual Sorvall ST 40 Sorvall ST 40R ...
Страница 6: ......
Страница 10: ...Tables iv Sorvall ST 40 Thermo Fisher Scientific ...
Страница 12: ...Figures vi Sorvall ST 40 Thermo Fisher Scientific ...
Страница 20: ...xiv Sorvall ST 40 Thermo Fisher Scientific ...
Страница 28: ...Introduction and Description Rotor selection 1 8 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 36: ...Before use Storage 2 8 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 37: ...Thermo Fisher Scientific Sorvall ST 40 3 1 Chapter 3 Control panel Control panel on page 3 2 Keys on page 3 4 ...
Страница 58: ...Operation Switching the centrifuge off 4 18 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 72: ...Maintenance and care Service of Thermo Fisher Scientific 6 12 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 80: ...Troubleshooting When to contact customer service 7 8 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 88: ...Warranty B 2 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 90: ...Index 2 Sorvall ST 40 Thermo Fisher Scientific Transporting the centrifuge 2 3 W WEEE Conformity 1 1 ...
Страница 93: ......