Maintenance and care
Autoclaving
6-10
Sorvall
ST 40
Thermo Fisher Scientific
Autoclaving
Rotors and adapters can be autoclaved at 121 °C.
The maximum permissible autoclave cycle is 20 minutes at 121 °C.
Clean the rotor before autoclaving and rinse it with distilled water.
Remove all accessories (tubes, adapters) from the rotor. Place the rotor
on a flat surface.
NOTE:
No chemical additives are permitted in the steam.
CAUTION
Never exceed the permitted temperature and duration when autoclaving.
If the rotor shows signs of corrosion or wear, it must be replaced!
Содержание Sorvall ST40
Страница 1: ...50119928 1 Thermo Fisher Scientific Sorvall ST 40 Sorvall ST 40R Instruction manual ...
Страница 2: ......
Страница 3: ...PN 50119928 1 Print 03 09 Instruction manual Sorvall ST 40 Sorvall ST 40R ...
Страница 6: ......
Страница 10: ...Tables iv Sorvall ST 40 Thermo Fisher Scientific ...
Страница 12: ...Figures vi Sorvall ST 40 Thermo Fisher Scientific ...
Страница 20: ...xiv Sorvall ST 40 Thermo Fisher Scientific ...
Страница 28: ...Introduction and Description Rotor selection 1 8 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 36: ...Before use Storage 2 8 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 37: ...Thermo Fisher Scientific Sorvall ST 40 3 1 Chapter 3 Control panel Control panel on page 3 2 Keys on page 3 4 ...
Страница 58: ...Operation Switching the centrifuge off 4 18 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 72: ...Maintenance and care Service of Thermo Fisher Scientific 6 12 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 80: ...Troubleshooting When to contact customer service 7 8 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 88: ...Warranty B 2 Sorvall ST 40 Thermo Fisher Scientific ...
Страница 90: ...Index 2 Sorvall ST 40 Thermo Fisher Scientific Transporting the centrifuge 2 3 W WEEE Conformity 1 1 ...
Страница 93: ......