28.1°C
Temperature Display
Shut Down
•
Press
to stop the bath heating.
Note
Pressing and holding
for three seconds stops the heating and powers down the
controller display. Press
to turn the display back on.
•
For bath shut down place the circuit protector located on the rear of the bath to the
O
position.
•
To disconnect the bath remove the electrical plug from its power source.
With the bath running, and after 30 seconds, if
none of the keys are pressed, the controller will
display only the reservoir fluid temperature and
any active feature. Press any key to return to the
previously displayed screen.
Changing the Real Time Adjustment (CAL RTA) requires a calibrated reference thermometer. If the
displayed temperature does not accurately reflect the actual temperature in the reservoir an RTA
value is required.
Note
The RTA applies only to the selected preset.
As an example, if the temperature is stabilized and displaying 20°C but a calibrated reference
thermometer reads 20.5°C, set the RTA to -0.5°C. After you enter an RTA value allow the display
to stabilize before verifying the bath temperature.
Press
to return to the home screen.
The changed RTA will not be displayed on the home screen until that preset is selected using the
Selecting a Preset
procedure.
Precision
|
4-7
Thermo Scientific
Operation
|
Chapter 4
Содержание Precision GP 02
Страница 24: ...DANGER WARNING CAUTION I II III NEC 100 80 40 C 50 50 O 60 C BG...
Страница 27: ...DANGER WARNING CAUTION I II III 100 80 C 40 C 50 50 O off 60 C EL...
Страница 45: ...2 4 Precision Thermo Scientific Chapter 2 General Information...
Страница 61: ...4 8 Precision Thermo Scientific Chapter 4 Operation...
Страница 63: ...5 2 Precision Thermo Scientific Chapter 5 Preventive Maintenance...
Страница 66: ......