Chapter 8
Data Transfer and Software Updates
Thermo Scientific
Orion Versa Star Pro Meter User Manual
109
Meter Calibration Data Example
Mode
Format Example
pH
Meter Type, Meter Serial Number, Meter Software Revision, Channel Number, Electrode
ID, Module Serial Number, Method Number, Calibration Date, Calibration Time, User ID,
Total Calibration Points,
(repeated for each calibration point) Calibration Point Number, pH Unit, pH Value, mV
Unit, mV Value, Temperature Value, Temperature Unit, Calibration Type,
(repeated for each point-to-point segment) Segment Number, Slope Value, Slope Unit,
(repeated for each point-to-point segment) Segment Number, Offset Value, Offset Unit,
Average Slope Value, Average Slope Unit, Calibration Log Number
Example (5 point calibration): VERSA STAR, V00000, 2.73, Ch-1, SENSOR-01,
VA12345, M-101, 15-03-14, 1:30:30 PM, USER-01, 5, 1, pH, 1.678, mV, 290.4, 25.0, C
(MAN), Auto, 2, pH, 4.01, mV, 174.3, 25.0, C (MAN), Auto, 3, pH, 6.997, mV, 0.0, 25.0, C
(MAN), Auto, 4, pH, 10.013, mV, -173.7, 25.0, C (MAN), Auto, 5, pH, 12.46, mV, -289.5,
25.0, C (MAN), Auto, 1, -49.8, mV, 2, -58.4, mV, 3, -57.6, mV, 4, -47.3, mV, 1, 25.4, mV,
2, -0.2, mV, 3, -0.2, mV, 4, -31.1, mV, 90.0, %, 101 <CR>
RmV
Meter Type, Meter Serial Number, Meter Software Revision, Channel Number, Electrode
ID, Module Serial Number, Method Number, Calibration Date, Calibration Time, User ID,
Relative mV Value, Relative mV Unit, Raw mV Value, Raw mV Unit, Temperature Value,
Temperature Unit, Relative mV Offset Value, Relative mV Offset Unit, Calibration Log
Number
Example: VERSA STAR, V00000, 2.73, Ch-1, SENSOR-01, VA12345, M-101, 15-03-14,
1:30:30 PM, USER-01, 199.8, RmV, 199.7, mV, 25.0, C (MAN), 0.1, mV, 101 <CR>
ORP
Meter Type, Meter Serial Number, Meter Software Revision, Channel Number, Electrode
ID, Module Serial Number, Method Number, Calibration Date, Calibration Time, User ID,
ORP Value, ORP Unit, Raw mV Value, Raw mV Unit, Temperature Value, Temperature
Unit, ORP Offset Value, ORP Offset Unit, Calibration Log Number
Example: VERSA STAR, V00000, 2.73, Ch-1, SENSOR-01, VA12345, M-101, 15-03-14,
1:30:30 PM, USER-01, 419.5, RmV, 199.8, mV, 25.0, C (MAN), 219.7, mV, 101 <CR>
ISE
Meter Type, Meter Serial Number, Meter Software Revision, Channel Number, Electrode
ID, Electrode Type, Module Serial Number, Method Number, Calibration Date,
Calibration Time, User ID, Total Calibration Points,
(repeated for each calibration point) Calibration Point Number, Concentration Value,
Concentration Unit, mV Value, mV Unit, Temperature Value, Temperature Unit,
Calibration Type,
(repeated for each point-to-point segment) Segment Number, Slope Value, Slope Unit,
(repeated for each point-to-point segment) Segment Number, Offset Value, Offset Unit,
Average Slope Value, Average Slope Unit, Blank Function, Blank Value, Blank Unit,
Calibration Log Number
Example (5 point calibration): VERSA STAR, V00000, 2.73, Ch-2, SENSOR-02, F-,
VA12345, M-201, 15-03-14, 1:30:30 PM, USER-01, 5, 1, 0.1, ppm, 174.2, mV, 25.0, C
(MAN), Manual, 2, 1, ppm, 116.2, mV, 25.0, C (MAN), Manual, 3, 10, ppm, 58.0, mV,
25.0, C (MAN), Manual, 4, 100, ppm, 0.0, mV, 25.0, C (MAN), Manual, 5, 1000, ppm, -
57.8, mV, 25.0, C (MAN), Manual, 1, -58.0, mV/dec, 2, -58.2, mV/dec, 3, -58.0, mV/dec,
4, -57.8, mV/dec, 1, 116.2, mV, 2, 116.2, mV, 3, 116.0, mV, 4, 115.6, mV, -58.0, mV/dec,
0, 0.000, ppm, 201 <CR>