Chapter 5
Conductivity Measurement Module
Thermo Scientific
Orion Versa Star Pro Meter User Manual
81
Conductivity Calibration Editing
Once the f3 (Cal Done) key is pressed and the calibration is saved, press the f2 (Cal Edit) key
to edit the calibration data and fix individual points without a full recalibration. In the calibration
edit mode, press the
or
key to highlight a calibration point and press the f2 (Delete) key
to delete the point or press the f3 (Re-Measure) key to re-measure the point. Follow the
onscreen text prompts and instructions to edit the selected calibration point.
Orion Conductivity Standard Values from 15 °C to 30 °C
Refer to the
Appendix
section of this user manual for additional values from 0 to 50°C.
Temperature
(°C)
Orion
111.9mS/cm
Standard
Cat. No. 011005,
01100510
(mS/cm)
Orion
12.9mS/cm
Standard
Cat. No.
011006,
01100610
(mS/cm)
Orion
1413μS/cm
Standard
Cat. No.
011007,
01100710
(μS/cm)
Orion
147μS/cm
Standard
Cat. No.
01100910
(μS/cm)
Orion
100μS/cm
Standard
Cat. No.
011008
(μS/cm)
15
92.34
10.46
1145
119
81
16
94.24
10.69
1171
122
83
17
96.15
10.93
1198
125
85
18
98.08
11.16
1224
127
87
19
100.0
11.40
1251
130
88
20
102.0
11.64
1277
133
90
21
103.9
11.88
1304
136
92
22
105.9
12.12
1331
138
94
23
107.9
12.36
1358
141
96
24
109.9
12.61
1386
144
98
25
111.9
12.85
1413
147
100
26
113.9
13.10
1441
150
102
27
115.9
13.35
1468
153
104
28
117.9
13.59
1496
156
106
29
120.0
13.84
1524
159
108
30
122.0
14.09
1552
161
110