90
Orion Versa Star Pro Meter User Manual
Thermo Scientific
7
Meter Measurements
Measurement Overview
The type of module connected to each of the four channels on the Versa Star Pro meter will
define the measurement capability of that channel.
Versa Star Module
Measurement Capabilities
pH module
pH, mV, relative mV or ORP with temperature
pH/ISE module
pH, mV, relative mV, ORP or concentration (ISE) with temperature
pH/LogR module
pH, mV or ORP with temperature/LogR temperature
Conductivity module
Conductivity, TDS, salinity or resistivity with temperature
RDO/DO module
Dissolved oxygen as % saturation or concentration with temperature
When two or more modules are connected to the meter, pressing the channel key will change
the number and combination of channels displayed in the measurement mode. Each channel
can be displayed individually or two to four channels can be displayed simultaneously.
The information shown in the measurement mode can be customized using the
Instrument
Settings
setup menu. The channel number, measure mode, method, stability indicator,
temperature, secondary measurement parameter, calibration details, user ID, sample ID and
electrode serial number can each be shown or hidden. Hiding these items in the measurement
display will increase the font size of the main measurement value in the single channel display.
Use the Buzzer – Ready setting in the
Instrument Settings
setup menu to enable an audible
beep when a measurement becomes stable.
Chapter 7