Chapter 6
RDO/DO Measurement Module
86
Orion Versa Star Pro Meter User Manual
Thermo Scientific
RDO and DO Mode Menu
Main Menu
Settings
Details
Electrode SI No
----
Use the alphanumeric keypad to enter the serial
number or clear all to shown no value (----)
Sample ID
Off, Manual, Auto Increment
Use the popup keypad screen to enter an
alphanumeric manual ID or a numeric value as
the starting value for auto incremental ID
Measurement Unit
mg/L,
%
Set the displayed measurement units
Resolution
0.1, 0.01
1, 0.1
Set the resolution of the measurement value
Stability
Smart, Fast, Medium, Slow
Set when a measurement is recognized as
stable
Averaging
Off, Automatic Smart
Set averaging for faster measurement stability
Read Type
Auto-Read, Timed,
Single-Shot, Continuous
For Timed and Single-Shot, set the time in hours,
minutes, seconds using the arrow keys
Pressure Comp
Auto, Manual
775.0
Set the barometric pressure compensation
source as Auto to use the built-in barometer or
Manual to enter a value. Select the Auto option
to calibrate the barometer pressure reading
Salinity Correction
Auto, Manual
Set the salinity correction source as Auto to use
a conductivity module and probe to measure the
sample salinity or Manual to enter a value
Alarm
Limit Alarm
Off, On
Set high and low limit values to activate alarm
Cal Due Alarm Off, On
Set a calibration interval (hours) to activate alarm
Set Point Alarm Off, On
Set a base and offset value to activate alarm
Detailed information on these features is available in the
Appendix
section of this user manual.
RDO Electrode Information
The serial number, remaining cap life and date of manufacture for each RDO optical cap can be
viewed in the RDO Electrode Information menu when the RDO probe is connected to the meter
and recognized in the measurement mode.
Verify that the RDO probe is connected to the meter. In the measurement mode, press the
setup key. Press the
or
key to highlight Channel # (channel with the RDO probe) and
press the f3 (Select) key. Press the
or
key to highlight RDO Electrode Information and
press the f3 (Select) key.