Thermo Scientific

     

Multifuge X1R 

 

1-3

1

Introduction and description

Technical data

Technical data

The technical data of the Multifuge X1R  is listed in the following table. 

Table 1-1. 

Technical

 

Data

 

Multifuge X1R 

Feature

Environmental conditions

Multifuge X1R 

-Use in interior spaces

-Altitudes of up to 2,000

 

m above sea level

-

M

ax. relative humidity 80

 

% up to 31 °C; decreasing linearly up to 50

 

%

relative humidity at 40 °C.

Permissible ambient temperature 

+2 °C to +

3

5 °C

Overvoltage category

II

Pollution degree

2

Voltage

230 V

Heat dissipation

4778 BTU/h

IP

20

Run time

U

nlimited

Max speed n

max

15200 rpm (depending on the rotor)

Min speed n

min

300 rpm

Maximum RCF value at n

max

25830 x

 

Maximum kinetic energy

62.5 kJ

Noise level at maximum speed 

< 63 dB (A)

Temperature setting range

-10 °C to +40 °C

Dimensions

R

efrigerated

Height
Height with lid open
Width
Depth
Table top height

360 mm 
870 mm
623 mm 
605 mm
310 mm

Weight without rotor

91.5 kg

 

Refrigerant 

Quantity   Pressure  

GWP 

 CO2e 

Order No.

75004249

R-

134

A

0.

2

8 kg 

31 bar 

3922 

       

1.49 t 

Centrifuge

Thermo Scientific

 

Multifuge X1R 

Contains fluorinated greenhouse gases in a hermetically sealed system.

Refrigerant 

Содержание Multifuge X1R

Страница 1: ...Thermo Scientific Multifuge X1R Centrifuge Instruction Manual 50156623 c 07 2020 Visit us online to register your warranty thermofisher com labwarranty...

Страница 2: ......

Страница 3: ...Mains supply 1 4 Rotor selection 1 5 Before use 2 1 Before setting up 2 2 Transporting the centrifuge 2 2 Location 2 2 Aligning the centrifuge 2 3 Mains Connection 2 3 Storage 2 4 Control panel 3 1 Co...

Страница 4: ...6 6 1 Cleaning intervals 6 2 Cleaning 6 2 Disinfection 6 4 Decontamination 6 5 Autoclaving 6 5 Service of Thermo Fisher Scientific 6 6 Troubleshooting Chapter 7 7 1 Mechanical emergency lid release 7...

Страница 5: ...s instruction manual will result in the expiration of the seller s warranty Scope of supply If any parts are missing please contact your nearest Thermo Fisher Scientific representative Article number...

Страница 6: ...liance with standards directives and trade association safety regulations and proper instruction of the user The safety regulations contain the following basic recommendations Maintain a radius of at...

Страница 7: ...he instructions on the Auto Lock in section Rotor installation on page 4 2 Use only with rotors which have been loaded properly Follow the instructions given in the rotor manual Never overload the rot...

Страница 8: ...es Do not centrifuge toxic or radioactive materials or any pathogenic micro organisms without suitable safety precautions When centrifuging microbiological samples from the Risk Group II according to...

Страница 9: ...d description Contents Characteristics of the Thermo Scientific Multifuge X1R on page 1 2 Technical data on page 1 3 Directives standards and guidelines on page 1 4 Functions and features on page 1 5...

Страница 10: ...try mode These settings can be changed even while the centrifuge is running With the help of the key you can also centrifuge a sample for just a few seconds if called for The Multifuge X1R is equipped...

Страница 11: ...tegory II Pollution degree 2 Voltage 230 V Heat dissipation 4778 BTU h IP 20 Run time Unlimited Max speed nmax 15200 rpm depending on the rotor Min speed nmin 300 rpm Maximum RCF value at nmax 25830 x...

Страница 12: ...hout carbon brushes Drive Easy to clean keypad and display surface Keys and display Controls Microprocessor controlled Internal memory The most recent data is saved Functions RCF selection temperature...

Страница 13: ...can be found in the corresponding rotor operating manuals For more information visit our website at http www thermofisher com centrifuge TX 400 75003181 with round buckets 75003655 TX 200 75003658 wi...

Страница 14: ...1 6 Multifuge X1R Thermo Scientific 1 Introduction and description...

Страница 15: ...Multifuge X1R 2 1 2 Before use Contents Before setting up on page 2 2 Transporting the centrifuge on page 2 2 Location on page 2 2 Aligning the centrifuge on page 2 3 Mains Connection on page 2 3 Sto...

Страница 16: ...gure 2 1 Lifting the centrifuge at both sides The centrifuge can be damaged by impacts Transport the centrifuge upright and if at all possible in its packaging Location The centrifuge should only be o...

Страница 17: ...ff the power supply switch on the back press 0 2 Plug the centrifuge into grounded electrical sockets only 3 Check whether the cable complies with the safety standards of your country 4 Make sure that...

Страница 18: ...entific Storage Before storing the centrifuge and the accessories it must be cleaned and if necessary disinfected and decontaminated Store the centrifuge in a clean dust free location Be sure to place...

Страница 19: ...Thermo Scientific Multifuge X1R 3 1 3 Control panel Contents Control panel on page 3 2 Keys on page 3 2...

Страница 20: ...mple temperature Key Display contents Start Normal start of the centrifuge Stop End run manually Open lid Automatic release possible only when device is switched on Emergency release see Mechanical em...

Страница 21: ...4 2 Door opening on page 4 2 Rotor installation on page 4 2 Entering parameters on page 4 4 Centrifugation on page 4 10 Temperature adaptation during standstill on page 4 11 Short term centrifugation...

Страница 22: ...lay shows The top line shows the message lid open The bottom line shows the values for the last run Door opening key c Press the WARNING Do not reach into the crack between the lid and the housing The...

Страница 23: ...he rotor is properly installed by lifting it slightly on the handle If the rotor can be pulled up then it must be reclamped to the centrifuge spindle CAUTION Unapproved or incorrectly combined accesso...

Страница 24: ...twice in order to open the braking curve selection menu The display shows the message Set deceleration Preselecting Speed RCF key 1 Press the Be sure to check all seals before starting any aerosol ti...

Страница 25: ...s a multiple of the force of gravity g It is a unitless numerical value which is used to compare the separation or sedimentation capacity of various devices since it is independent of the type of devi...

Страница 26: ...anually with the Run time mode Use the key to set the run time counter start AT START When AT START is lit the run time will be counted right from pressing the AT SPEED When AT SPEED is lit the run ti...

Страница 27: ...Depending on the display mode setting the value shown corresponds either to the sample temperature or to the inside chamber temperature Press the key to toggle between the two modes 2 Enter the desir...

Страница 28: ...ress the key repeatedly until the bucket being used is displayed Your entry will be automatically confirmed if you do not press any key for 5 seconds Bucket radius Having entered the bucket code you m...

Страница 29: ...memory key to load any of the programs 1 5 Press the key in order to load any of the programs 1 99 by means of the numeric pad Enter the program number and confirm with 3 You may want to enter a passc...

Страница 30: ...en the lid if the centrifuge is running Imbalance indicator If a load is imbalanced this will be indicated at speed higher than approx 300 rpm by the message Imbalance load The run will terminate Chec...

Страница 31: ...he centrifuge lid 2 Grab the rotor handle with both hands and press against the green Auto Lock key At the same time pull the rotor directly upwards with both hands and remove it from the centrifuge s...

Страница 32: ...you and the samples Figure 4 3 Auto Lock lid for aerosol tight rotors Turning off the centrifuge 1 To turn off the centrifuge put the mains switch to 0 CAUTION Rotors supplied with a lid for aerosol t...

Страница 33: ...Thermo Scientific Multifuge X1R 5 1 5 System menu System menu on page 5 2 Description on page 5 3 Contents...

Страница 34: ...5 Nederlands 6 7 STOP ENTER LCD powersave START no 0 yes 1 ENTER ENTER No Program passcodes 0 0547V011 1 Software and NVRAM identification example display 4641V05 1 ENTER STOP ENTER Set lid passcode E...

Страница 35: ...stem menu hold down any of the keys when switching the centrifuge on Press the key in order to navigate through the system menu Press the key to quit the system menu Language In order to change the la...

Страница 36: ...after the run Otherwise press 0 no Press key to confirm Software ID Here you fine the current software version Press key to confirm Cycle counter Here you fine the current numbers of cycles Press key...

Страница 37: ...ifuge X1R 6 1 6 Maintenance and care Contents Cleaning intervals on page 6 2 Cleaning on page 6 2 Disinfection on page 6 4 Decontamination on page 6 5 Autoclaving on page 6 5 Service of Thermo Fisher...

Страница 38: ...drying boxes are used the temperature must never exceed 50 C since higher temperatures could damage the material and shorten the lifetime of the parts Use only disinfectants with a pH of 6 8 Dry alumi...

Страница 39: ...ning with a cloth or in a warm air cabinet at a maximum temperature of 50 C 9 After cleaning treat the entire surface of aluminum parts with corrosion protection oil 70009824 Also treat the cavities w...

Страница 40: ...8 Rinse the rotor and accessories thoroughly with water 9 Dispose of the disinfectant according to the applicable guidelines 10 Dry all the rotors and accessories after cleaning with a cloth or in a w...

Страница 41: ...otors and accessories after cleaning with a cloth or in a warm air cabinet at a maximum temperature of 50 C 11 After cleaning treat the entire surface of aluminum parts with corrosion protection oil 7...

Страница 42: ...d lock and the safety system the rotor the fixation of the rotor and the drive shaft Thermo Fisher Scientific offers inspection and service contracts for this work Any necessary repairs are performed...

Страница 43: ...Thermo Scientific Multifuge X1R 7 1 7 Troubleshooting Contents Mechanical emergency lid release on page 7 2 Troubleshooting by user on page 7 3 When to contact customer service on page 7 6...

Страница 44: ...usual Proceed as follows 1 Make sure the rotor has stopped view port in the lid 2 Pull out the power supply plug 3 On the left side of the housing is one white plastic plugs which you can pry out of...

Страница 45: ...rotor The centrifuge cannot be operated The run does not start or the centrifuge runs down without being braked Rotor code is not in the rotor table Is it permitted to use the rotor currently mounted...

Страница 46: ...art the centrifuge If an error message appears again inform your service technician E 13 The centrifuge cannot be operated The run does not start or the centrifuge runs down without being braked The c...

Страница 47: ...E 33 The centrifuge cannot be operated The run does not start or the centrifuge runs down without being braked Overpressure in the refrigeration unit Clean the air inlet for the condenser Restart the...

Страница 48: ...o of your device This information can be found on the back near the inlet for the power supply cable To identify the software version proceed as follows 1 Hold down any of the keys and then switch on...

Страница 49: ...U S S Formic Acid 100 S M U U U S M U U S S U S U S U Ammonium Acetate S S U S S S S S S S S S S U S S S S S S S S S S Ammonium Carbonate M S U S S S S S S S S S S S U U S S S S S S M S S S Ammonium H...

Страница 50: ...M S U U S U M U U M M M U S S S S U U Ferric Chloride U U S M S M S S S S M U S S Acetic Acid Glacial S S U U S S U M S U S U U U U U M S U M U S U U S U Acetic Acid 5 S S M S S S M S S S S S M S S S...

Страница 51: ...S S S S M S M U S S Kerosene S S S S S S U S M U S U M M S M M M S S U S S U S Sodium Chloride 10 S S S S S S S S S S S S S S S S S S M S Sodium Chloride sat d U S U S S S S S S S S S S S S S M S Carb...

Страница 52: ...cid 10 U U S U U S M M M U M S M M M S U S S Perchloric Acid 70 U U U U U S U M U U M U U U M M U M S U U S U S Phenol 5 U S U S M M S U M U U S U M S M S U U S U M M M S Phenol 50 U S U S U M S U M U...

Страница 53: ...U U U U U U U S U U S Trisodium Phosphate S M S S S S S S S Tris Buffer neutral pH U S S S S S S S S S S S S S S S S S S S S S S S S S Triton X 100 S S S S S S S S S S S S S S S S S S S S S S S S S Ur...

Страница 54: ...terials under the stress of centrifugation When in doubt we recommend pretesting sample lots Key Satisfactory S M Moderate attack may be satisfactory for use in centrifuge depending on length of expos...

Страница 55: ...ing 4 2 E Emergency door release 7 2 Entering parameters 4 4 Explanation of RCF value 4 5 F Failure message 7 3 Features 1 5 I Imbalance indicator 4 10 K Keys 3 2 L Location 2 2 M Mains Connection 2 3...

Страница 56: ...zone 2 2 Select program 4 9 Short term centrifugation 4 11 Starting centrifuge program 4 10 Stopping the centrifugation program 4 10 Storage 2 4 Switch on centrifuge 4 2 T Technical Data 1 3 Temperat...

Страница 57: ...Thermo Scientific Multifuge X1R I 3 I Index...

Страница 58: ...es Please consult your local sales representative for details Shown pictures within the manual are examples and may differ considering the set parameters and language Pictures of the user interface wi...

Отзывы: