6
Maintenance and care
6-2
Multifuge X1R
Thermo Scientific
Cleaning intervals
For the sake of personal, environmental, and material protection, it is your duty to clean and if
necessary, disinfect the centrifuge on a regular basis.
Cleaning
When cleaning centrifuge mind the following:
•
Use warm water with a neutral solvent.
•
Never use caustic cleaning agents such as soap suds, phosphoric acid, bleaching solutions or
scrubbing powder.
•
Rinse the cavities out thoroughly.
•
Use a soft brush without metal bristles to remove stubborn residue.
•
Afterwards rinse with distilled water.
•
Place the rotors on a plastic grate with their cavities pointing down.
•
If drying boxes are used, the temperature must never exceed 50 °C, since higher temperatures could
damage the material and shorten the lifetime of the parts.
•
Use only disinfectants with a pH of 6-8.
•
Dry aluminum parts off with a soft cloth.
CAUTION
During maintenance operations, always disconnect the centrifuge from the
power.
Maintenance
Recommended interval
E
very month or when polluted
Cabinet / Housing
D
aily or when polluted
Rotor chamber
D
aily or when polluted
Rotor
D
aily or when polluted
Accessories
O
nce per month
Cabinet
E
very six months by a service technician
Ventilation holes
CAUTION
Refrain from using any other cleaning or decontamination procedure than those
recommended here, if you are not entirely sure that the intended procedure is safe for the
equipment.
Use only approved cleansers.
If in doubt, contact Thermo Fisher Scientific.
Содержание Multifuge X1R
Страница 2: ......
Страница 14: ...1 6 Multifuge X1R Thermo Scientific 1 Introduction and description...
Страница 19: ...Thermo Scientific Multifuge X1R 3 1 3 Control panel Contents Control panel on page 3 2 Keys on page 3 2...
Страница 33: ...Thermo Scientific Multifuge X1R 5 1 5 System menu System menu on page 5 2 Description on page 5 3 Contents...
Страница 57: ...Thermo Scientific Multifuge X1R I 3 I Index...