Sample loading guidelines
•
Use the recommended total loading volume for each gel type
. Do not load more than
recommended amount of DNA sample or ladder per well.
•
Load deionized water into all empty wells
.
•
Keep all sample volumes uniform
. If you do not have enough samples to load all the wells of
the gel, load an identical volume of deionized water into any empty wells. Prepare your
samples by adding E-Gel
™
1X Sample Loading Buffer or deionized water to the required
amount of DNA to bring the total required sample volume.
•
Avoid introducing bubbles while loading. Bubbles can cause band distortion.
Gel type
Total loading volume
E-Gel
™
EX
20 μL
E-Gel
™
with SYBR
™
Safe
E-Gel
™
EX Double Comb
E-Gel
™
Double Comb with SYBR
™
Safe
E-Gel
™
Go!
10 μL
E-Gel
™
CloneWell II
25 μL
E-Gel
™
SizeSelect II
25 μL
E-Gel
™
NGS
20 μL
Load samples
1.
Load prepared samples. Keep all sample
volumes uniform.
2.
Load prepared DNA ladder.
3.
Load 1X E-Gel Sample Loading Buffer
or
deionized water in all empty
wells.
4.
Run gels within 1 minute after loading
samples.