32
E-Gel
™
Power Snap Electrophoresis System User Guide
NGS library size selection reference
Library Size
Target
library
peak
Run time to
reference
line
Input
sample
amount
Stop the run and collect your
sample when…
Schematic
view
Ion S5™ XL System
600-base-read
680 bp 17.5-19 min
500 ng
600 bp band has just completely
entered the top edge of the
collection well
50-100 ng
700 bp band has just completely
entered the top edge of the
collection well
Ion PGM™ System
400-base-read
480 bp
14–20 min
500 ng
500 bp band is at the top of the
exposed agarose area
50–100 ng
500 bp band has just entered the
top edge of the collection well
300-base-read
390 bp
13–16 min
500 ng
400 bp band is at the middle of
the exposed agarose area
50–100 ng
500 bp band is at the top of the
exposed agarose area
200-base-read
330 bp
12–14 min
500 ng
350 bp band is at the top of the
exposed agarose area
50–100 ng
350 bp band has just completely
entered the top edge of the
collection well
100-base-read
200 bp 11–12.5 min
500 ng
200 bp band is in the middle of
the collection well
50–100 ng
200 bp band is in the middle of
the collection well
Ion Proton™ System
200-base-read
270 bp
12–14 min
500 ng
300 bp band is at the top of the
exposed agarose area
50–100 ng
300 bp band is at the middle of
the exposed agarose area
150-base-read
220 bp 11–14.5 min
500 ng
250 bp band is at the middle of
the exposed agarose area
50–100 ng
250 bp band is at the middle of
the exposed agarose area