18
E-Gel
™
Power Snap Electrophoresis System User Guide
Run the gel
1.
Press
Set up run
to start E-Gel
™
protocol selection.
2.
Select
the E-Gel
™
protocol corresponding to your gel
type.
Use the up/down arrows to navigate through the
menu.
3.
(
Optional
) For recurring experiments, select the last
used protocol.
Gel Type
Recommended
program
Default run
time
Maximum run time
E-Gel
™
EX Agarose Gel, 1% and 2%
E-Gel EX 1-2%
10 min
20 min
E-Gel
™
EX Agarose Gel, 4%
E-Gel EX 4%
15 min
20 min
E-Gel
™
EX Double Comb Agarose Gels,
1% and 2%
E-Gel EX 1-2%
5 min
8 min
E-Gel
™
Agarose Gel with SYBR
™
Safe,
1%, 2%
E-Gel 0.8-2%
26 min
40 min
E-Gel
™
Agarose Gel with SYBR
™
Safe,
4%
E-Gel 4%
30 min
40 min
E-Gel
™
Double Comb SYBR
™
Safe
Agarose Gels, 1% and 2%
E-Gel Double Comb
13 min
18 min
E-Gel
™
CloneWell
™
II Agarose Gel,
0.8%
CloneWell 0.8%
12 min
40 min
E-Gel
™
SizeSelect
™
II Agarose Gel, 2%
SizeSelect 2%
8 min
20 min
E-Gel
™
NGS
™
Agarose Gel, 0.8%
E-Gel 0.8-2%
26 min
32 min
E-Gel
™
Go! Agarose Gel, 1% and 2%
E-Gel Go! 1-2%
15 min
30 min
Reverse protocol for:
E-Gel
™
CloneWell
™
II Agarose Gel
E-Gel
™
SizeSelect
™
II Agarose Gel
Reverse E-Gel
2 min
3 min