E-Gel
™
Power Snap Electrophoresis System User Guide
55
Installing
the
instrument
The product may be installed only under the conditions and in the positions specified by Thermo
Fisher Scientific.
Following are the required operating position and conditions:
•
Do not place the product in an area where it will be subject to vibration.
•
Do not place the product on surfaces, vehicles, cabinets or tables that for reasons of weight or
stability are unsuitable for this purpose.
•
Do not place the product on heat-generating surface or near heat emitting devices such
equipment racks or heaters. Verify that there is sufficient clearance between the product and
any other system that may exhaust warm air.
•
The product's ventilation should not be obstructed. If proper ventilation is not provided it can
result in electric shock, fire and/or serious personal injury or death.
•
The product is for indoor use only
•
Use only with suitably rated mains supply cord (having 3 conductors min. 16 AWG or 1.5
mm
2
, min. 300V, Harmonized Type for Europe and UL Listed/CSA Certified for North
America, with molded plug rated min. 10A).
•
A tolerance of ±10 % shall apply to the nominal input voltage and ±3 Hz to the nominal
frequency, overvoltage category 2.
•
Maximum operating altitude 2000 m asl, Maximum transport altitude
4500 m asl.
Electromagnetic compatibility (EMC) standards
Class A notice
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications.