Note:
Attaching the top tabs will help keep the BPC
properly oriented inside the tank. Figure 32 shows the
top front BPC tab attached to the tab holder.
Figure 33. Inserting the RTD.
Figure 34. RTD in a probe clip.
• Place a probe clip on the probe support hanger and
place the RTD so it is resting in the hook (Figure 34).
Figure 35. Foam probe in holder.
29. Clamp the black foam probe holder to the outer
support container, behind the top left tab holder, if
facing the front of the S.U.B. Insert the foam probe into
the foam probe holder on the outer support container
(Figure 35).
Figure 31. Placing black top tab
holders on the top front of tank.
Figure 32. Black top tab holder
with BPC tab attached.
24. Gently pull at the bottom corners of the BPC until the
sides of the BPC are straight.
Note:
You may need
to wait for the BPC to fill with more air before this is
possible.
25. While the BPC continues filling with air, remove all
bubble wrap and polybags from the line sets.
26. Remove the polybag from the drain line set, position
the line clamp as close as possible to the BPC port,
and then close the clamp. Use a cable tie around the
clamp to ensure it does not open.
27. Remove the plastic insert located in the thermowell, if
present.
28. Insert the resistance temperature detector (RTD) into
the thermowell (Figure 33) using the steps below.
Note:
Due to the shorter thermowell on DynaDrive BPCs, the
15.24 cm (6 in.) RTD is the ONLY compatible size.
DO
NOT
use larger RTDs, which will damage the BPC.
• Place a small amount of glycerol (0.5 mL) in the
thermowell to aid in heat transfer. The glycerol also
serves as a lubricant and aids in insertion.
• The sensor should be inserted until the base of the
RTD meets the mouth of the thermowell.
• Secure by twisting the luer lock collar, if provided.
CAUTION:
• The foam probe must be placed into the foam
probe holder to prevent the foam sensor inside the
BPC from becoming entangled with the drive train.
• The 500 L foam probe holder must be removed
prior to raising or lowering the motor lift to avoid
damaging the BPC.
30. Optional: Connect a pressure sensor to the CPC
aseptic connector at the top of the BPC. Then connect
the appropriate pressure transducer cable to the third-
party controller.