4
Setting up
Installing ECM rinse liquid bag
22
Thermo Scientific
Figure 18.
Reference filling solution bag installed
12. Close the ECM unit cover.
Note
Calibrate the tests and run quality controls after working with the ECM unit.
Installing ECM rinse liquid bag
To install the ECM rinse liquid bag:
1. Connect the tube of the rinse liquid bag to the tube coming from the ECM dispensing
pump.
2. Hang the bag in its place.
3. Select F5 > Actions >
Add ECM rinse liquid
to fill the tube with liquid.
Testing ECM electrodes
To test the electrodes:
1. Run start-up procedure as described in Operation Manual.
2. Run and accept a calibration.
3. Review the calibrations of each electrode.
For more information about calibration, refer to Operation Manual.
Note
If problems occur, check the connections of electrodes and rinse liquid bag.