Thermo Scientific
Configuring analyzer settings
Analyzer settings are configured by the service engineer after installing the analyzer. To configure
the settings, select F5 > Configuration >
Analyzer.
Figure 20.
Analyzer settings
Table 11.
Analyzer configuration parameters
Parameter
Description
Analyzer type
Define the analyzer type.
Application area
Shows the application area.
Vendor
Shows the vendor of the analyzer.
Tick interval (sec)
Shows the length of the tick in seconds. Tick is a basic measurement unit for the analyzer.
Additional module
Define the additional module used in the analyzer. If no additional module is used, select
None
.
The default value is set by the manufacturer.
Max water blank SD (A)
Define the maximum allowed limit for water blank standard deviation (SD).
Max allowable absorbance (A) Define the maximum allowed absorbance value. If the value exceeds the defined value, an error
message is displayed in result reports.
Reagent disk cooling
Define whether the reagent disk is cooled or not.
LIS connection in use
Define whether the LIS connection is in use or not.
Содержание Gallery Plus
Страница 1: ......
Страница 6: ...vi Thermo Scientific Figure 2 Safety labeling inside analyzer ...
Страница 12: ... this page was intentionally left blank ...
Страница 18: ... this page was intentionally left blank ...
Страница 26: ... this page was intentionally left blank ...
Страница 44: ...4 Setting up Setting up 26 Thermo Scientific 7 Connect the workstation s power cable ...
Страница 47: ...A Thermo Scientific 29 Removing transport casing ...
Страница 48: ... this page was intentionally left blank ...
Страница 52: ... this page was intentionally left blank ...
Страница 56: ... this page was intentionally left blank ...