4
Setting up
Setting up
Thermo Scientific
19
Table 10.
Order of ECM electrodes from rod to probe
Order
Electrode
Color code
Color of
connecting wire
1
White Assembly A (not connected)
-
-
2
Reference electrode
Brown
Brown
3
pH electrode
White
White
4
EC Assembly
Transparent
Black
5
Spacer slice
Transparent
-
6
White Assembly B
Transparent
Transparent
Figure 16.
Connecting wires and electrodes
Note
Calibrate the tests and run quality controls after working with the ECM unit.
Содержание Gallery Plus
Страница 1: ......
Страница 6: ...vi Thermo Scientific Figure 2 Safety labeling inside analyzer ...
Страница 12: ... this page was intentionally left blank ...
Страница 18: ... this page was intentionally left blank ...
Страница 26: ... this page was intentionally left blank ...
Страница 44: ...4 Setting up Setting up 26 Thermo Scientific 7 Connect the workstation s power cable ...
Страница 47: ...A Thermo Scientific 29 Removing transport casing ...
Страница 48: ... this page was intentionally left blank ...
Страница 52: ... this page was intentionally left blank ...
Страница 56: ... this page was intentionally left blank ...