64
EVOS
™
M5000 Imaging System User Guide
Calibrate white balance
Calibrate white
balance
Calibrate White Balance calibrates color channel lighting.
1.
Go to the
Settings
Visuals
tab, then click
White Balance Calibration
to open
the White Balance Calibration tool.
2.
Select a 20X or higher objective, then set the phase ring to
Brightfield
using the
Phase annuli selector (page 7).
3.
Remove any samples from the X-Y stage, then center the stage using the stage
positioning knobs (page 7).
4.
Click
Start
to begin the automatic white balance calibration procedure.