EVOS
™
M5000 Imaging System User Guide
63
Calibrate objective
magnification
1.
Go to the
Settings
Objective Selection and Calibration
,
click the
Action
button next to the newly installed objective, then select
Calibrate
to launch
the Objective Calibration tool.
2.
Mount the EVOS
™
Calibration Slide in the vessel holder, select the objective
you want to calibrate, then click
Calibrate Objective
.
3.
Follow the onscreen instructions to calibrate the objective magnification for the
selected objective. For more information, see “Calibrate the objectives”, page 76.
Delete objectives
After removing an objective from the objective turret, unassign the objective from
the corrsponding turret position.
1.
Go to the
Settings
Objective Selection and Calibration
,
click the
Action
button next to the turret position from which you have removed the
objective, then select
Delete
.