5.
Press the
Calibrate
function key.
When Measure Standards = Yes, you will be prompted to measure each
standard in turn.
When Measure Standards = No, the calibration will be calculated from
the standards that have been entered.
6.
Save the calibrated method to the instrument Library or to disk.
A previously calibrated method can be loaded from the instrument
Library or from disk.
When a valid calibration is present, pressing Run causes the absorbance
of the sample to be measured at each of the wavelengths specified in the
method. The concentration of each component is then calculated from
the calibration.
7.
Press the
View Results
function key to transfer to the MCA Results
page.
The up/down arrow keys can be used to view previous/following
results.
8.
Press the
Save Results
function key to save results to the instrument
Library or to disk.
Biological tests
Follow these steps to run Biological tests:
The menu enables you to move between the four major applications.
To run any of the biological tests available, highlight
Biological Tests
on the
Main Menu and press
Enter
.
56 Evolution 300 and Evolution 600 User Guide