Installing Smart Accessories
Positive location and automatic alignment of the accessories is achieved
using pins A and B and the spring loaded toggle as described for manual
accessories. In addition, data communication is achieved via the serial
D connector underneath the front of the accessory, which plugs into the
corresponding socket on the accessory. A tapered guide rod is provided
beside the connector to assist in aligning the connector and socket during
installation.
1.
Open the main lid of the sample compartment and remove any
accessory that may be present in the sample (front) position.
2.
Remove the front wall of the sample compartment by sliding it
upward in its grooves.
3.
Insert the tapered guide pin on the bottom front of the accessory
into the corresponding guide pin hole located next to the
D connector (female).
Do not press down to completely seat the accessory.
4.
Position the accessory so that alignment holes A and B are aligned
with pins A and B on the instrument and press down gently.
22 Evolution 300 and Evolution 600 User Guide