Parameter
Function
Autoprint Results
Toggled On – test setup, data and results are
automatically printed at the end of each run. This
option uses the currently selected printer.
Toggled Off – no automatic printout. User must Print
to obtain a printout if desired.
AutoSave Results
Toggled On – test results are automatically saved
after each experiment in a user named file stored in
either the internal instrument memory or on a disk.
When On, the user must enter a name deed and
start number.
Toggled Off – data is not automatically saved. User
must select Save Data for each set of data if they
want to save the data.
Clock
Allows user to set the time and date.
Date Format
Selects the desired date format.
Printer
Selects the active printer.
Language
Selects the appropriate display language.
Sound
Toggled On – any error is indicated by an on screen
message and audible beep.
Toggled Off – on screen messages are the only
indication of an error.
History File
Toggles On – significant events (e.g., recalibration)
are automatically recorded in the History File.
Toggled Off – no events can be recorded in the
History File.
Live Status
Toggled On – the display of the live reading at the
top right hand corner of the screen includes current
wavelength and data.
Toggled Off – the display of the live reading at the
top right hand corner of the screen includes only the
current wavelength.
Print -> Screen.bmp
Toggled On – allows the user to create a bitmap of
the current screen contents to be saved to a disk
when the Print button is pressed.
Toggled Off – printing function is directed to the
installed printer.