Installation
28
|
Dionex IC Pure system
Thermo Scientific
Chapter 7
Connecting of a pretreatment cartridge (optional)
Connecting an external 20 L tank (optional)
CAUTION
When using a pretreatment cartridge, the initial pressure at the connections for the IC
unit must be reset. Contact the Service department at Thermo Scientific for this.
Always ensure that the internal 5 L or external 20 L tank does not drain completely, as this may
damage or render the optional pretreatment cartridge unusable.
When using a pretreatment cartridge, drain the internal 5 L or external 20 L tank to a level of 30%
and then fill up the tank manually (see “Draining the internal 5 L and external 20 L tank” ).
Step
Action
Figure
1
a.
Pull the tube jumper out of its connections by
pressing on the quick-connectors.
b.
Proceed as described in Step 3a on page 25
and install the pretreatment cartridge in the
front section of the unit. When you hear an
audible click, the quick-connectors have been
inserted correctly. The arrangement of the
connections prevents any incorrect/switched
installation.
Note
Keep the tube jumper in a safe place for later use,
e.g., in the disinfection procedure.
Tube jumper
Quick-connectors
a.)
Concentrate
Note
The external 20 L tank (50146376) will not be use the additional tank indicator on the IC
Pure system directly. After filling the tank manually and connecting the control cable to the IC Pure
system, the additional tank indicator is ready for control the level in the external 20L tank after
switching on.
Note
When operating the unit without a pretreatment cartridge, manually fill the 20 L tank with
pretreated ASTM Type II water. The tank should only be filled with potable tap water if the
pretreatment cartridge is used in the Dionex IC Pure system.
CAUTION
Always ensure that the external 20 L tank does not drain completely, otherwise an air
lock between the external 20 L tank and Dionex IC Pure system can be formed and this can affect the
lifetime of the pretreatment cartridge.
Note
Ensure that the Dionex IC Pure system has been switched off and the internal 5 L tank is
empty (see draining 5 L tank section). In addition, the 20 L tank should be located directly next to the
Dionex IC Pure system.
Содержание dionex ic pure water
Страница 10: ......
Страница 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Страница 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Страница 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Страница 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Страница 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Страница 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Страница 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Страница 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Страница 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Страница 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Страница 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Страница 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Страница 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Страница 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Страница 127: ...Thermo Scientific IC Pure 117 ...