Set the interval pump time
The interval pump time is the amount of time the pump is working to recirculate water in the system.
The standard setting is 5 minutes of pump recirculation for every 30 minutes that the system stands
still during “Interval mode“. The majority of systems do not need this setting to be changed.
Basic setting: 5 minutes
Setting range: 1 - 30 min.
After entering the OEM menu, press the “Menu“ button four times. The display shows:
Set the interval rinse time
In this menu the rinse interval time can be set. When the system operates in the Interval mode, the
system will rinse water to drain for 0.5 seconds every 30 minutes.
Basic setting: 0.5 sec.
Setting range: 0.1 - 2 sec.
After entering the OEM menu, press the “Menu“ button five times. The display shows:
Set the real-time clock
The real-time clock can be set in this menu.
Basic setting: The actual date
Setting range: 1 - 12 month; 1 - 31 day; 0 - 24 h; 0 - 60 min.
After entering the OEM menu, press the “Menu“ button six times. The display shows:
Note
The system is factory adjusted in the system control. In this section it is not necessary to be
carried out changes of value for operating with an IC system.
Содержание dionex ic pure water
Страница 10: ......
Страница 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Страница 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Страница 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Страница 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Страница 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Страница 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Страница 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Страница 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Страница 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Страница 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Страница 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Страница 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Страница 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Страница 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Страница 127: ...Thermo Scientific IC Pure 117 ...