Maintenance
80
|
Dionex IC Pure system
Thermo Scientific
Chapter 13
Replacing the pretreament cartridge (optional)
7
a.
Keep pressing the “Menu“ button until you see “Ser. No.:
Cartridge“ in the display.
b.
Enter the new serial number. Use the “NONSTOP and
INTERVAL“ button to change the numeric values and the
“UV“ button to move to the next value.
c.
After you have entered the serial number, confirm your entry
by pressing “Enter“. The new serial number for the ultrapure
cartridge is then stored. You can only use a ultrapure
cartridge serial number one time.
8
Step
Action
Figure
Note
For this action you must have a Level 1 code. Refer to
section “Entering a Code number” on page 71 to read where to find
this code and how to enter it.
a.)
b.)
c.)
Note
Discharge the first 5 L.
Note
The pretreatment cartridge should only be replaced when the maximum limit for the
feedwater has been exceeded, when you cannot extract any water from the sampling tap, or when the
replacement interval is due.
Содержание dionex ic pure water
Страница 10: ......
Страница 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Страница 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Страница 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Страница 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Страница 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Страница 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Страница 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Страница 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Страница 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Страница 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Страница 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Страница 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Страница 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Страница 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Страница 127: ...Thermo Scientific IC Pure 117 ...