Thermo Scientific
Dionex IC Pure system |
99
Troubleshooting
Note
Contact the service department if you cannot rectify this error.
Error
Cause
Remedy
The system does not start
No supply of power
Provide power
Dispensing not possible
Feedwater supply line for
external tank filling station
closed
Feedwater and rinse water
connections are mixed up
Feedwater pressure < 100 kPa
(14.5 psi)
Sterile filter blocked
Pretreatment cartridge blocked
Open the feedwater supply line
Replace connections
Increase feedwater pressure
Replace the sterile filter (see
“Replacing the 0.2 µm final filter”
on page 92
Install new pretreatment cartridge
Resistance < 18.2 M
·cm
Cartridge is exhausted
Temperature compensation
deactivated, recalibration
required
Check feedwater quality
Activating temperature
compensation ('TC' should appear
in the display) Contact the service
organization for calibration
Control panel non responsive
Microprocessor locked up
PCB error
Unplug the mains plug for 5
seconds.
Contact the Thermo Fisher
Scientific service department
Water leak
Leaky tubing connection
Feedwater pressure > 600 kPa
(87 psi) (external tank filling
station only)
Check and seal the tubing
connection
Install a pressure reducer
Contact the Thermo Fisher
Scientific service department
Содержание dionex ic pure water
Страница 10: ......
Страница 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Страница 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Страница 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Страница 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Страница 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Страница 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Страница 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Страница 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Страница 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Страница 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Страница 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Страница 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Страница 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Страница 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Страница 127: ...Thermo Scientific IC Pure 117 ...