3-18
DB-070 E
RadEye AB100
Pi/Ff 04.05.2021
Count rate
Measuring
time
Additional
value
ADF
Linear
ratemeter
0…<8 cps
10 s
80
x
8…<32 cps 5s
40
x
>32 cps
1s
0
x
All count
rates
Preset time
constant
(1..60s)
0
x
Thus the actual alarm threshold adjusts itself automatically to the
existing background radiation level. The actual setting of the
alarm threshold 1 can be observed at the bar graph threshold
mark. At power on of the instrument, the background Ratemeter
starts with its preset maximum background value, reaching its
final value after about 2 minutes.
Using the menu option
Alarm
αβ
, the exact value of the alarm
threshold currently valid at the respective background can be
displayed.
On acknowledgment of an alarm, the background Ratemeter is set
to the actual count rate display. Accordingly the alarm threshold
is increased.
There will be no further alarms at background conditions and
normal sensitivity will be established again automatically after
about two minutes.