RadEye AB100
DB-070 E
12-9
Pi/Ff 04.05.2021
V 1.75
•
Scaler mode can be controlled remotely
•
Rounding error in measuring unit “cpm” and “dpm” fixed
•
Menu item “Display α/αβ” was disabled. Now it is visi-
ble again.
•
New function: “Alarm acknowledge” enable/disable. If
Alarm acknowledge is disabled the acoustic and vibration
alarm cannot be muted.
•
New function: Set alarm threshold to 3xbackground. Af-
ter background measurement the user is asked if the alarm
threshold for beta threshold should be set to
3xbackground count rate.
•
LED flash and a double beep at the end of a scaler meas-
urement.
•
Data set number is increased at the start of a scaler meas-
urement.
•
Lowest settable voltage: 100V. Before the voltage could
be set to 0V
•
New menu items “Display contrast” and “Language” in
sub menu “Settings”
•
Feature "beta window in activity mode" removed
•
A classic linear ratemeter filter can be selected instead of
ADF.
•
The ratemeter can be set to the actual value by a long
press on the “Menu/◄” button.