8-6
DB-070 E
RadEye AB100
Pi/Ff 04.05.2021
By pressing the [OK] button the user is prompted to either con-
firm or negate the question “Accept parameter?” By pressing the
[No] button you will quit the plateau measurement parameters
entry mode and any modifications made will be cancelled. To
adopt the new plateau measurement parameters, press [Yes] and
the plateau parameter entry mode is quit automatically.
8.1.3
Starting a plateau measurement
Once the user has pressed the “Start measurement” button, the
Program checks whether a stored plateau has been loaded and the
displayed plateau parameters are up-to-date. In the event that the
displayed parameters are no longer the current ones, the current
plateau parameters will be displayed and the user is prompted to
start the measurement process. Then an “identification text” needs
to be entered. This identification text is shown in the title box and
is saved together with the plateau. It serves to later identify the
plateau measurement.
Figure 8-5
Plateau identification text
After completing the entry of the identification text, the value of
the high voltage is first set to its start value and a there is a 6 sec-
ond wait during the stabilization period. Then the first individual
measurement is started.
Содержание DB-070 E
Страница 1: ...Operating Instructions DB 070 E RadEye AB100 Multi Purpose Survey Meter ...
Страница 2: ......
Страница 12: ......
Страница 42: ......
Страница 56: ......
Страница 80: ...7 18 DB 070 E RadEye AB100 Pi Ff 04 05 2021 Figure 7 15 Logbook ...
Страница 92: ......
Страница 98: ......
Страница 103: ...RadEye AB100 DB 070 E 11 3 Pi Ff 04 05 2021 ...
Страница 104: ......
Страница 114: ......
Страница 116: ......
Страница 117: ......