Thermo Scientific
7-1
7
Excess Temperature Option(“B” Model)
The Excess Temperature Option, when installed, provides an additional,
independent temperature control system to help protect products from excess
temperatures.
Read this section carefully before using this option.
Control Display
When the Excess Temperature Controller is first turned on, it displays only the
excess temperature setpoint in the bottom display. Press and release the
SET/ENT button to show the duration time of the last excess temperature incident.
(See
“Exceeded Temperature Duration Timer Section”
Press and release the SET/ENT button once again to show the peak temperature
measured for the last excess temperature incident.
Pressing and releasing the SET/ENT button once again shows the current
temperature measured by the controller in the top display; this may differ slightly
from the main temperature controller. The value shown in the bottom display is the
current excess temperature setpoint.
Pressing and releasing the SET/ENT button again cycles back to the first display.
Excess Temperature Option Features
1. Exceeded Temperature Duration Timer
The Exceeded Temperature Duration Timer measures the time that the
setpoint is exceeded (and power to the heater interrupted) until the hysteresis
value is reached as the chamber temperature cools. This time indicates when
the chamber temperature exceeded the Excess Temperature setpoint.
2. Peak Exceeded Temperature
The Peak Exceeded Temperature is the highest temperature measured by the
Excess Temperature Controller.
Содержание BF51841BC
Страница 1: ......
Страница 6: ...List of Figures ii Thermo Scientific ...
Страница 18: ...Safety Notes Standards and Directives 1 12 Thermo Scientific ...
Страница 22: ...Unpacking Unpacking 3 2 Thermo Scientific ...
Страница 28: ...Installation Shelf Installation in Box Furnaces 4 6 Thermo Scientific ...
Страница 38: ...Correction of Temperature Overshoot 6 8 Thermo Scientific ...
Страница 42: ...Excess Temperature Option B Model Excess Temperature Controller Operation 7 4 Thermo Scientific ...
Страница 48: ...Communication Option Addresses for Multiple Controllers 8 6 Thermo Scientific ...
Страница 55: ...Maintenance Door Insulation Replacement Thermo Scientific 9 7 Figure 11 Door Insulation Replacement ...
Страница 57: ...Maintenance Right Hand Door Conversion Thermo Scientific 9 9 Figure 12 Door Hinge Sheet 2 2 ...
Страница 63: ...Replacement Parts Thermo Scientific 11 3 Figure 13 Wiring Diagram ...
Страница 64: ...Replacement Parts 11 4 Thermo Scientific ...
Страница 65: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable ...
Страница 66: ...Error Codes 12 2 Thermo Scientific ...
Страница 68: ...WEEE Compliance 13 2 Thermo Scientific ...
Страница 69: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable ...
Страница 70: ...Spare Parts and Accessories 14 2 Thermo Scientific ...
Страница 72: ...Device Log 15 2 Thermo Scientific ...