Auto Tuning the Controller
6-6
Thermo Scientific
5. Press the arrow buttons to select the new Up Ramp Rate value, in °C per
minute or °F per minute, or “OFF”. Press and release SET/ENT to register the
value change.
6. Press and release SET/ENT until the dnr parameter is on the upper display.
7. Press the arrow buttons to select the new Down Ramp Rate value, in °C per
minute or °F per minute, or “OFF”. Press and release SET/ENT to register the
value change.
8. Press and hold SET/ENT for three seconds to exit the Setup Parameters
menu.
9. The new Ramp Rates are now effective.
10. Follow the instructions in
Section “Normal Controller Operation”
to reset the
temperature setpoint.
To view the target setpoint during the ramp rate, press and release an arrow
button. The lower display will show the setpoint next to the selected target
setpoint. Return to the ramping setpoint display by pressing and releasing the
other arrow button.
Auto Tuning the Controller
Factory settings are provided for general purposes, but your process can be
enhanced through the auto tune feature. For a given process temperature and
product load, auto tuning maximizes the performance of the chamber by operating
with the quickest response and minimal temperature overshoot.
To auto tune the controller.
1. Load the chamber with materials that have the same mass and thermal
characteristics as a typical product load.
2. Operate the chamber to the process temperature.
3. Press and hold SET/ENT for 3 seconds to display the A1 parameter of the
Operating Parameter menu.
4. Press and release SET/ENT to show the At parameter.
5. Press and release the arrow buttons show “ON” in the lower display.
6. Press SET/ENT once to enter the auto tune mode and exit the Operating
Parameters menu.
The ramp rate begins when the SET/ENT button is pressed after target
setpoint is selected. The setpoint display on the controller will show the
changing setpoint at the selected ramp rate.
Содержание BF51841BC
Страница 1: ......
Страница 6: ...List of Figures ii Thermo Scientific ...
Страница 18: ...Safety Notes Standards and Directives 1 12 Thermo Scientific ...
Страница 22: ...Unpacking Unpacking 3 2 Thermo Scientific ...
Страница 28: ...Installation Shelf Installation in Box Furnaces 4 6 Thermo Scientific ...
Страница 38: ...Correction of Temperature Overshoot 6 8 Thermo Scientific ...
Страница 42: ...Excess Temperature Option B Model Excess Temperature Controller Operation 7 4 Thermo Scientific ...
Страница 48: ...Communication Option Addresses for Multiple Controllers 8 6 Thermo Scientific ...
Страница 55: ...Maintenance Door Insulation Replacement Thermo Scientific 9 7 Figure 11 Door Insulation Replacement ...
Страница 57: ...Maintenance Right Hand Door Conversion Thermo Scientific 9 9 Figure 12 Door Hinge Sheet 2 2 ...
Страница 63: ...Replacement Parts Thermo Scientific 11 3 Figure 13 Wiring Diagram ...
Страница 64: ...Replacement Parts 11 4 Thermo Scientific ...
Страница 65: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable ...
Страница 66: ...Error Codes 12 2 Thermo Scientific ...
Страница 68: ...WEEE Compliance 13 2 Thermo Scientific ...
Страница 69: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable ...
Страница 70: ...Spare Parts and Accessories 14 2 Thermo Scientific ...
Страница 72: ...Device Log 15 2 Thermo Scientific ...